Introduction of HPC600 Handheld Six-Channel Hybrid Particle Measuring Counter
The HPC600 handheld laser particle counter
is the latest innovation in the demanding application of particle distribution
measurements. It is useful in measuring particle distributions in ultra-clean
environments by its single particle counting ability as well as in indoor air
quality applications. The instrument consists of a handheld monitor with a main
base unit that allows users to conduct sampling with the portable handheld
monitor while easily expanding to multiple functionalities provided by the base
unit. These extended functions include real-time data remote sampling,
printing, data downloading, software upgrades, battery charging, etc. In
addition to USB interfaces, the RJ45 interface allows users to conduct remote
sampling away from the sampling location via Ethernet/Internet.
The HAL-HPC600 is the world’s first of
its kind hybrid handheld optical particle counter. The instrument is truly a
breakthrough for a new generation of particle counters that combine the
traditional handheld instruments with functionalities of portable instruments.
It is in compliance with the international standards (JIS B 9925:1997 and ISO14644-1)
and CE certification. It supports both metric and English systems. All of its
key components are made in the USA, Germany or Japan. It features high
sensitivity, multiple functional capabilities, ease of use and reliability for
extremely sensitive environmental measurements and advanced applications.
Features
of HPC600 Handheld Six-Channel Hybrid Particle Measuring Counter
?All 6 channels continuously adjustable with 0.1 micron step
resolution
?Combines both handheld and portable functionalities with a built-in
mini-printer
?USB and RJ45 interfaces for remote real-time sampling and data
download
?Long-distance remote sensing via Ethernet
?Up to 6000 internal data memory
?Excess-count-limit warnings
?High precision, digital, external temperature and humidity sensors
Durable keypads with a large blue LCD
display
Applications of HPC600 Handheld Six-Channel Hybrid Particle Measuring Counter
?Clean environment monitoring
?Indoor air quality
?Test/Check filter seal and efficiency
?Trace contamination source
?Analysis of particle size distribution
Specifications of HPC600 Handheld Six-Channel Hybrid Particle Measuring Counter
Light Source
|
Laser diode (>100,000 hours MTBF)
|
Sensitivity
|
0.3μm
|
Size Range
|
0.3 ~ 25μm
|
Channels
|
User configurable on any channel with 0.1μm resolution (six default
channel settings: 0.3, 0.5, 0.7, 1.0, 2.0, 5.0μm)
|
Counting Efficiency
|
50±20% @ 0.3μm
100±10% @ 0.5μm
|
Coincidence Loss
|
<5% @ 70,000 particles/liter
|
Zero Count
|
<1 count per 5 minutes
|
Flow Rate
|
2.83 L /min (0.1cfm)
|
Sampling Time
|
User defined: (up to 59m59s) and auto repeat (up to 99 times)
|
Count Limit Warning
|
FED STD 209E (Class 1 ~ 300,000)
or ISO 14644-1 (Class 2 ~ 9)
|
Measurement
|
Single/repeat/Average
|
Error Indication
|
Count limit, loss of laser power, out of flow calibration
(>±5%),
insufficient battery
|
Interface
|
USB,
|
Internal Memory
|
6000 measurement data (1000 sets)
|
Power
|
Removable, rechargeable Lithium battery (7.4V/2800mAh) or
AC Adapter (input:AC100~240V,
output: DC9V/1.5A)
|
Max. Operating Time
|
Continuous
operation > 5 hours with Lithium battery
|
Dimensions
|
Handheld: 185 (H) × 93 (W) ×46 (D) mm
Base unit: 152 (Dia.)× 97(H) mm
|
Weight
|
< 800 grams (including battery)
|
Environmental Conditions
|
Operating: 5 ~ 45°C, < 90%RH
Storage: -20 ~ 50°C, < 90%RH
|
Accessories
|
Main base unit, AC
adaptor, iso-kinetic probe, USB data cable, remote sampling and data download
software (CD), zero-count/purge filter, traceable calibration certificate
|
Images of HPC600 Handheld Six-Channel Hybrid Particle Measuring Counter